Products
| HS Code | 847439 |
| Product Name | Honeywell HPLC Grade Isopropyl Alcohol (IPA) |
| Chemical Name | 2-Propanol |
| Cas Number | 67-63-0 |
| Molecular Formula | C3H8O |
| Molecular Weight | 60.10 g/mol |
| Purity | >99.9% |
| Grade | HPLC |
| Appearance | Colorless clear liquid |
| Density | 0.785 g/mL at 25°C |
| Boiling Point | 82.5°C |
| Flash Point | 11.7°C (closed cup) |
| Refractive Index | 1.377 at 20°C |
| Solubility | Miscible with water |
| Water Content | ≤0.02% |
As an accredited Honeywell HPLC Grade Isopropyl Alcohol (IPA), 2 Propanol>99.9%,Isopropanol Solution factory, we enforce strict quality protocols—every batch undergoes rigorous testing to ensure consistent efficacy and safety standards.
| Packing | Honeywell HPLC Grade Isopropanol (IPA), >99.9%, is packaged in a 4-liter glass bottle with a secure, airtight cap. |
| Container Loading (20′ FCL) | 20′ FCL loaded with Honeywell HPLC-grade IPA (>99.9% isopropanol), drums secured and labeled, no co-loading. |
| Shipping | This chemical ships as a hazardous material (Class 3 flammable liquid) in compliant, sealed containers. It must be transported ground or air, adhering to IATA/IMDG regulations. Proper labeling, UN1219 identification, and secure upright packaging prevent leakage and ensure safe delivery to laboratories and industrial facilities. |
| Storage | Store Honeywell HPLC-grade isopropanol in a tightly sealed, original container in a cool, dry, well-ventilated area away from heat, sparks, and open flames. Keep away from strong oxidizers and incompatible materials. Avoid exposure to moisture and contamination. Ensure the container remains upright and properly labeled, with appropriate flammable storage practices. |
| Shelf Life | Stable for 3-5 years if unopened and stored in original container, away from heat and light. Ensure tight sealing. |
The 99.9 % assay 2-propanol stream described here is used in reversed-phase HPLC and UHPLC laboratories where weak baseline drift at low ultraviolet wavelengths cannot be tolerated. During reversed-phase gradient elution, isopropanol acts as a stronger elution solvent than acetonitrile or methanol for highly retained lipids, peptides, and synthetic polymers. At 210 nm detection wavelengths, trace carbonyl-containing impurities in reagent-grade isopropanol produce excessive background absorbance, so the chromatographic grade must report UV transmittance against a water reference at 210 nm, 220 nm, and 230 nm. The solvent lot acceptance window for water is typically below 0.05 % w/w, and non-volatile residue is controlled below 2 ppm because non-volatile material deposits on the electrospray ion source and the detector flow cell. In normal-phase separations of phospholipids, 5–30 vol% 2-propanol in hexane is used to elute phosphatidylcholine, sphingomyelin, and lysophospholipid classes from silica columns. Because isopropanol has a viscosity of approximately 2.04 mPa·s at 25 °C, higher than methanol and acetonitrile, binary pump pressure limits often require column temperature settings between 40 °C and 60 °C when the isopropanol volume fraction exceeds 30 %. Column equilibration times must be extended when the mobile phase contains isopropanol because the solvent reorders the bonded phase more slowly than acetonitrile. System suitability under USP <621> includes repeated retention time and resolution checks across a gradient sequence, and residual solvent compliance is interpreted under ICH Q3C when the method serves pharmaceutical quality control. Terminal products from this application include validated analytical methods for drug substance release, lipid impurity profiles in parenteral nutrition, and polymer molecular weight distributions by size-exclusion chromatography.
Single-wafer cleaning tools in 300 mm logic and memory fabs use heated isopropanol vapour in Marangoni drying after aqueous HF and SC1/SC2 cleaning sequences. The drying mechanism depends on the lower surface tension of isopropanol relative to ultra-pure water; when IPA vapour condenses on the wafer surface, surface-tension gradients pull water out of high-aspect-ratio trenches and vias. Semiconductor-grade 2-propanol used in this operation is specified under SEMI C35 and controlled for alkali metals, transition metals, and anionic species that cause mobile ion contamination in gate oxide layers. Vapour dryers typically operate at 80–90 °C with nitrogen carrier gas and a DI water rinse immediately upstream. The explosion-proof exhaust system is mandatory because isopropanol vapour forms flammable mixtures in air between 2.0 % and 12.7 % by volume. Published batch data for this specific configuration is limited to supplier-supplied ICP-MS certificates, which report metals in the low µg/kg range; critical elements include sodium, potassium, calcium, iron, copper, and zinc. Failure modes observed on manufacturing lines include threshold-voltage shifts from alkali contamination and increased defect density from drying watermarks when the IPA/water contact zone is unstable. The high-purity alcohol is used undiluted in vapour dryers, while wafer backside cleaning may use a 70:30 v/v isopropanol-water mixture for organic residue removal before plasma deposition. Terminal products include advanced node CMOS wafers, 3D NAND memory, MEMS structures, and photomasks.
In API finishing suites, the final rinse after aqueous clean-in-place cycles uses high-purity 2-propanol to displace water from stainless-steel agitators, transfer lines, and filter housings. Water left on equipment surfaces after CIP can hydrolyse moisture-sensitive intermediates and increase bioburden during extended hold time. A two-step sequence is common: a 70:30 v/v isopropanol-water solution is used as a sanitising wipe or recirculation rinse, followed by undiluted high-purity isopropanol as a drying rinse. The solvent is handled under nitrogen pressure in closed equipment to reduce peroxide formation; recovery is by atmospheric distillation with bottoms temperature not exceeding 80 °C. In wet granulation, 2-propanol can replace ethanol as a binder solvent for moisture-sensitive actives. Typical granulation solvent loading varies from 20 % to 50 % w/w of dry powder mass depending on API particle size and binder type. Granulation endpoint is determined by torque or power draw on high-shear mixers, and residual solvent levels are checked by headspace GC according to USP <467>. Terminal products include film-coated tablets, hard gelatin capsules, and sterile injectables requiring ICH Q3C residual solvent control.
| Reference | Parameter | Limit or Requirement |
|---|---|---|
| ICH Q3C | Class 3 PDE for isopropanol | 50 mg/day |
| ICH Q3C | Concentration limit for Class 3 solvents | 5000 ppm |
| USP <467> | Residual solvent measurement method | Headspace GC |
| 21 CFR 211.67 | Equipment cleaning procedure requirement | Written procedures |
Sterile filling lines in isolator-based aseptic processing rely on 70 % v/v isopropanol and 30 % v/v water-for-injection as a routine surface disinfectant. The water content slows evaporation and allows penetration through microbial cell walls. The solution is pre-filtered through a 0.45 µm membrane and sterile-filtered through a 0.2 µm membrane into autoclaved stainless-steel or HDPE containers. In Grade A transfer isolators, contact time is maintained between 30 s and 60 s before transfer of stoppers, vials, and filling needles. The high-purity 2-propanol lot used for this purpose must have low subvisible particulate counts and low non-volatile residue; residue accumulation on isolator gloves and windows disrupts visibility and can shed particles. EU GMP Annex 1 requires a rotating sporicidal agent because isopropanol has limited activity against bacterial endospores; this limitation is controlled by alternating with hydrogen peroxide or peracetic acid. Terminal products include ready-to-fill syringes, lyophilised vials, and ophthalmic suspensions.
Hair spray systems based on octylacrylamide/acrylates/butylaminoethyl methacrylate copolymers require a low-water alcohol stream to avoid resin precipitation during aerosol filling. In pump spray and aerosol formulations, high-purity isopropanol is added at 5–15 wt% as a co-solvent with ethanol and n-butanol to modify solution viscosity and evaporation rate. The solvent blend influences spray particle size distribution; a slower-evaporating alcohol reduces dry spot formation on hair fibres. Batch filling under nitrogen pressure prevents atmospheric water ingress that would cause turbidity in clear cans. The formulation is filtered through 5 µm or 10 µm polypropylene bag filters before cold filling to remove undissolved polymer gels. Under Regulation (EC) No 1223/2009, the final product must have a cosmetic product safety report; isopropanol is not a prohibited substance in Annex II and is used within general safety limits. Published data for the specific HSP resin-isopropanol solubility limit is limited; laboratory screening at the target alcohol blend is required before scale-up. Terminal products include aerosol hair sprays, nail polish removers, and hair styling gels.
Solventborne flexographic inks printed on corona-treated BOPP or LDPE use solvent blends in which 10–30 wt% 2-propanol functions as a co-solvent for nitrocellulose or polyurethane resins. The low surface tension of 2-propanol improves wetting of non-polar film surfaces and assists pigment wetting in bead-mill dispersion. In a typical manufacturing sequence, pigment is dispersed in a bead mill at 3000–5000 rpm with a solvent-to-pigment ratio between 1.5:1 and 4:1. High-purity 2-propanol is added during letdown to adjust final viscosity to a printing viscosity of 18–25 s measured by a Zahn cup No. 2. Non-volatile residue in the high-purity alcohol is controlled because residue accumulates on anilox cells and changes ink transfer density. For thin-film conformal coating thinners used in printed circuit board assembly, residual metals in solvent must meet customer-defined limits under IPC J-STD-001 to avoid ionic contamination. Compliance is anchored to ASTM D2369 for volatile organic content and REACH substance restrictions; food-contact printed films require that solvent residues after drying meet EU 10/2011 overall migration limits. Terminal products include surface-printed snack packaging, adhesive lamination primers, and conformal coating thinners.
Competitive Honeywell HPLC Grade Isopropyl Alcohol (IPA), 2 Propanol>99.9%,Isopropanol Solution prices that fit your budget—flexible terms and customized quotes for every order.
For samples, pricing, or more information, please contact us at +8618136850665 or mail to sales4@ascent-chem.com.
We will respond to you as soon as possible.
Tel: +8618136850665
Email: sales4@ascent-chem.com
Flexible payment, competitive price, premium service - Inquire now!
Honeywell HPLC Grade Isopropyl Alcohol (IPA), 2-Propanol 99.9%, Isopropanol Solution is supplied under the Burdick & Jackson high-purity solvent line for reversed-phase and normal-phase liquid chromatography, sample preparation, and instrument wash applications. The product is identified by CAS 67-63-0 and EINECS 200-661-7; the 4 L amber-glass presentation, commonly ordered as AH323-4, is used in analytical laboratories where solvent purity must be matched to detector baseline requirements. Gas chromatographic assay release data typically report ≥99.9% 2-propanol. Water content is controlled by Karl Fischer titration, evaporation residue by gravimetric determination, and UV absorbance by spectrophotometry in 1 cm quartz cells against a water reference. These release controls distinguish the product from commodity or ACS-grade isopropanol, which may meet general laboratory purity limits but do not consistently address detector baseline stability, piston seal wear, or column contamination risk. The substance has a molecular weight of 60.10 g/mol, a density of approximately 0.786 g/cm³ at 20 °C, a boiling point of 82.5 °C at 101.3 kPa, a closed-cup flash point of 12 °C, and a vapour pressure of 4.4 kPa at 20 °C.
In reversed-phase gradient separations with photodiode-array detection, the solvent must remain optically transparent at the monitored wavelength. IPA has a UV cutoff near 205 nm; below this threshold, the solvent itself absorbs strongly and reduces the dynamic range available for analyte quantitation. Lot release documentation for the product addresses this through absorbance ceilings at 210 nm, 220 nm, 230 nm, and 260 nm. Typical acceptance limits are 1.00 AU at 210 nm, 0.50 AU at 220 nm, 0.20 AU at 230 nm, and 0.05 AU at 260 nm in a 1 cm cell. These limits do not guarantee that every detector system will be noise-free at 210 nm; detector lamp intensity, flow cell pathlength, and gradient mixing efficiency also contribute to baseline drift. For methods operating below 210 nm, the solvent absorbance profile should be checked against the specific detector and column because small lot-to-lot shifts in trace UV-absorbing impurities may be magnified by gradient compression. In practice, the ≥99.9% GC assay is less important for spectroscopic behaviour than the non-solvent impurity classes; concentration of aldehydes, ketones, and unsaturated oxygenated species at part-per-million levels is controlled by the UV absorbance release test. The lower UV transmittance of IPA relative to acetonitrile or methanol requires a higher reference photomultiplier gain or longer integration time when using UV detection at 210 nm to 220 nm. For electrospray ionisation LC-MS, the product may be used as an organic modifier, but dedicated LC-MS-grade solvents are separately controlled for metal ions and plasticizers; published data for this specific HPLC-grade IPA configuration is limited, and laboratories requiring trace metal values should request lot-specific certificates rather than infer equivalence.
| Parameter | Test method | Typical release limit |
|---|---|---|
| Assay | Gas chromatography, flame ionisation detection | ≥99.9% |
| Water | Karl Fischer titration; ASTM E203-08 | ≤0.05% |
| Evaporation residue | Gravimetric; ASTM D1353-13 | ≤5 mg/L |
| UV absorbance at 210 nm | Spectrophotometry; USP 857 | ≤1.00 AU |
| UV absorbance at 220 nm | Spectrophotometry; USP 857 | ≤0.50 AU |
| UV absorbance at 230 nm | Spectrophotometry; USP 857 | ≤0.20 AU |
| UV absorbance at 260 nm | Spectrophotometry; USP 857 | ≤0.05 AU |
| Acidity/alkalinity | Titration | ≤0.0002 meq/g |
| Density at 20 °C | ASTM D4052-22 | 0.784–0.787 g/cm³ |
The water and UV absorbance limits in the release matrix are not redundant. A solvent can pass a 99.9% GC assay while still failing a low-wavelength absorbance test because UV-active impurities at 0.01% or less may be sufficient to raise background absorbance. This is a key difference from bulk chemical purity grades. In gradient HPLC, the UV background is not constant across the solvent programme. When the IPA fraction increases from 5% to 95%, any solvent impurity is concentrated in the mobile-phase front, exposing downstream detector cells to transient absorbance excursions. The release test therefore uses fixed-wavelength measurements in a 1 cm cell as a surrogate for detector response. For critical impurity profiling, a blank gradient should be executed after each new lot, and the data system should be configured to compare blank baseline slope between the previous lot and the new lot. If the blank baseline slope at 210 nm changes by more than the method's integration threshold, the lot should not be used without verifying detector lamp age and flow cell cleanliness.
Normal-phase retention time reproducibility is sensitive to water in the mobile phase. At the product's water limit of ≤0.05%, silica and cyano columns equilibrate more predictably than with technical-grade isopropanol, which may contain several tenths of a percent of water. For example, a hexane/IPA gradient used for class separation of phospholipids or fat-soluble vitamins will shift retention times when water dissolved in IPA alters the activity of silica silanol sites. The Karl Fischer water value is therefore not merely a purity benchmark; it is part of the effective mobile-phase composition. Laboratories that replace fresh, tightly sealed bottles with partially used bottles that have been exposed to ambient humidity can exceed the water limit even if the original solvent met the certificate of analysis. A bottle opened at 60% relative humidity can absorb atmospheric water and gradually move retention times in normal-phase methods. For critical separations, the solvent system should be prepared fresh, and the IPA bottle should be blanketed with dry nitrogen if multiple acquisitions span days. In reversed-phase methods, the water content of an added organic modifier is less impactful, but the same headspace moisture uptake can influence gradient proportioning and biological sample reconstitution.
In high-pressure gradient pumps, the compressibility and viscosity of isopropanol differ from acetonitrile. IPA has a higher viscosity; at 20 °C, pure isopropanol is approximately 2.0 mPa·s, roughly double that of acetonitrile at 0.37 mPa·s. Isopropanol-water mixtures can generate higher backpressure than methanol-water or acetonitrile-water at the same flow rate and column temperature. The pressure rise is not a product defect but a physical property that must be accounted for when setting system upper-pressure limits. A method transferred from acetonitrile to IPA requires re-evaluation of the pump stroke volume, gradient proportioning valve timing, and seal wash flow. If the system is fitted with a 10 mm UV flow cell, the higher viscosity can also alter detector cell pressure; most analytical flow cells are rated for pressures of 6.9 bar to 10.3 bar, but backpressure from the column often provides a margin below the rated value. Users should monitor system pressure with a pressure transducer and set an alarm at 80% of the lowest-rated component.
In sample preparation, technical-grade isopropanol may introduce nonvolatile residue, phthalates, or trace carbonyl impurities that concentrate during nitrogen blowdown or solid-phase extraction elution. The HPLC-grade product's evaporation residue specification of ≤5 mg/L is intended to reduce the mass of nonvolatile material that would otherwise be co-concentrated with analytes. In methods using large solvent volumes—such as liquid-liquid extraction of environmental samples with 100 mL or more of extractant per sample—the residue contribution becomes analytically significant when the final extract is reconstituted in 1 mL to 2 mL of mobile phase. A solvent with 5 mg/L residue can contribute up to 0.5 mg of nonvolatile contamination per 100 mL extraction volume if no further cleanup is applied. The assay of ≥99.9% reduces co-eluting impurity peaks in gas and liquid chromatograms, but it does not eliminate the need for matrix blanks, because sample-derived interferences remain.
For instrument wash cycles, the product is used in aqueous/organic seal wash and needle wash solutions. The low residue limit protects high-pressure pumping systems from deposits on check valves and pistons; however, IPA is a stronger solvent for some polymeric seals than methanol or acetonitrile. Seal compatibility should be verified against the pump manufacturer's chemical resistance tables before long-term exposure. The product should not be used as a sole wash solvent for buffers that crystallize in pure organic solvents; a water-miscible intermediate wash is required to prevent salt precipitation in the pump head.
| Property | HPLC-grade IPA | Technical/commodity IPA | Chromatographic consequence |
|---|---|---|---|
| UV absorbance at 210 nm | Controlled by release test | Not specified | High background noise in gradient UV methods |
| Evaporation residue | ≤5 mg/L | May be higher and unspecified | Nonvolatile deposits on inlet liners, columns, and piston seals |
| Water | ≤0.05% | Not controlled to HPLC requirements | Normal-phase retention drift and sample moisture burden |
| Packaging filtration | Submicron-filtered solvent | Not guaranteed | Particulate scoring of pump pistons and column frits |
| Certificate of analysis | Lot-specific release data | May be limited to bulk assay | Traceability for regulatory method validation |
For LC-MS, the product can be used in the sample preparation stage, but dedicated LC-MS-grade solvents are controlled for sodium and potassium at low parts-per-billion levels because metal adducts suppress ionisation. The HPLC-grade product is not excluded from LC-MS use, but its certificate of analysis may not include metal ion data. Therefore, if the mass spectrometer is configured for negative-mode or low-signal analyses, the laboratory should not assume equivalence between the HPLC-grade and LC-MS-grade product. This is a product-differentiation boundary rather than a quality failure. For gradient LC-MS where the spray-chamber temperature exceeds 300 °C, the lower water and residue values still reduce mobile-phase-generated background ions.
The product remains flammable under normal ambient conditions, with a closed-cup flash point of 12 °C and a lower explosion limit of 2.0 vol% in air. Storage requires ventilated flammables cabinets, bonded grounding for bulk containers, and exclusion of ignition sources. The safety data sheet carries hazard classification for flammable liquid category 2, serious eye irritation category 2, and specific target organ toxicity single exposure category 3. Use is typically governed by local chemical hygiene plans and 29 CFR 1910.106 for US facilities or Directive 1999/92/EC for EU installations. These classifications are independent of chromatographic performance and must be incorporated into storage and handling systems.
Air exposure during bottle dispensing is not merely a flammability concern. Isopropanol can form organic peroxides on prolonged storage under oxygen and light, although at a slower rate than diethyl ether or tetrahydrofuran. Peroxide formation in secondary alcohols proceeds through free-radical autoxidation. After an induction period, alkyl hydroperoxides and hydrogen peroxide may accumulate; the induction period is shortened by UV light, transition metal ions, and oxygen partial pressure. A partially used 4 L bottle with a loose cap can develop measurable peroxide levels over months. Standard peroxide test strips with a detection limit of 0.5 mg/L are used for screening. Laboratories that retain partially used bottles for more than twelve months should implement peroxide screening; distillation or evaporation of aged material to dryness must be avoided because peroxide decomposition can be violent. The amber glass packaging reduces photochemical degradation, but it does not remove dissolved oxygen. For long-term stability, blanketing with nitrogen and storage in a certified flammable-liquid refrigerator or solvent cabinet is required by standard chemical hygiene practice.
The product is not a substitute for USP/NF-grade isopropanol in drug formulation, nor is it automatically equivalent to LC-MS-grade isopropanol. In regulated analytical methods, the user should verify that the certificate of analysis meets the specific pharmacopeial or compendial requirements for water, residue, UV absorbance, and assay. Method transfer from one production site to another should compare the lot release certificate against the method's established blank and baseline acceptance criteria, because solvent-related baseline artefacts may appear only in the gradient region where the proportion of IPA is highest. For methods requiring low peroxide background—such as derivatisation or oxidative-sensitive sample preparation—fresh lot inventory is preferable. Distillation of aged isopropanol for solvent purification should be performed with stirring and temperature monitoring; do not distil to a dry residue, and peroxide-negative verification should precede any thermal concentration step.